Description: New Gas Detector to check poisonal Gas w264291
A 100% Test in just 10 Seconds
Personnel working in
chemical plants protect themselves with personal
gas detectors from
gas exposure. Gas leaks are very unlikely to occur, so the
gas detector will always
display 0.
An instrument displaying zero can do so for two reasons:
a) The concentration is zero.
b) The instrument is defective.
For
good reasons a daily check of the instrument is mandatory. It would be too time consuming to perform a complete calibration every day. Local regulations allow calibration intervals of up to six months.
The daily functional test with the
gas generator makes sure that no instrument failure caused by abuse or wear and tear will go unnoticed. To perform a test it is sufficient to expose the
sensor to a short puff of
gas and see if the instrument gives an
alarm. To make this test as fast, easy and cheap as possible, no accurate
gas concentration is used, but approximately twice the value of the
alarm threshold.
The
quality criterion for the detector is not reaching a certain value, but the response time: A
good sensor is also a fast
sensor. So the
gas generator stops
gas generation after 10 seconds automatically. If the detector fails to
alarm within this time interval it is not safe to be used. Thus the
generator test reliably indicates the following potential errors:
- Defective
electronics- Alarm buzzer or
lamp defective
- Sensor defective
- Sensor too slow
- Sensor not sensitive enough
Compur Monitors introduces a new
gas generator model with the following features:
The flow of the built- in
pump is adjusted to perfectly simulate the operating conditions of the
gas detector.
The
gas generator itself consists of an electrochemical cell filled with a electrolyte paste. This cell will only generate
gas when it is activated by putting a
gas detector on the receptacle.