Description: 1. hermetic solid-state
optical system, stable performance with low contamination of dust and dampness, longer life time of hard membrane filters than soft ones. 2. Light split after the cuvette, corresponding every wavelength, there are independent
amplifier . Switch between different wavelength completed by
electric multiplexer, no mechanical movement, useful for bichromatic measurement. 3. Dedicated touch-input keyboard, optional Menu and instruction step by step. It can checking the status of the unit. 4. 240X128 dot matrix
LCD with back-lit, can
display ASC code and curve simultaneously, readable even if in the dim
light. 5. Show the real time. 6. 58mm thermal printer, high speed and lower noise, optional external
printer. 7. Measure methods: End point method, Fixed time method, Kinetic method and coagulation time method. 8. Light source: halogen lamp12V/ 20W. 9. Optical system: 340, 405, 505, 546, 578, 630nm, precision: +/-3nm. 10. Absorbance range:0.0000-2.5000ABS 11. After 10 minutes warm up, the drift<0.003Abs/hr 12. cuvette:10mm, 33ul, when the aspiration volume no less than 500ul, the carry over <1.0% 13. temperature control: 25,30,32,37 with 0.1more or less. 14. precision: when Abs.=0.5000,CV<0.1%; when Abs.=2.0000,CV<1%. 15. 96 different analysis can be set in. 16.
power supply: 110/220Vwith10% more or less, 50/60Hz, <80VA 17.RS232: standard serial port.
| Place of Origin: | China|shandong qingdao |