Product Description
ETCO2(End-Tidal Carbon Dioxide)technology for mainstream CO2 monitoring

PRODUCT FEATURES:

  • Ideal for non-intubated patients
  • Dual wavelength ,proprietary sample cell protects internal ndir components from contamination
  • Low sampling rate down to 50ml/min
  • Common connector allows easy exchange between mainstream and sidestream monitoring
  • Robust and long life pump reduces periodic maintenance
  • No calibration required
  • Private label option  TTL/RS232,LEMO 8 pins/DB9
  • Unique accessories & supplies for all patients
 
 
Transducer TypeSidestream CO2  Sensor
Sample Rate50 mL/min, ±10mL/min.
Principle of OperationNon-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
Initialization TimeCapnogram displayed in less than 4 seconds
full specifications within 2 minutes
CO2 Measure Range0 - 114 mmHg
0 - 15%
0 - 15.2kPa
CO2 Resolution0.1 mm Hg 0 to 59 mm Hg
0.25 mm Hg 60 to 114 mm Hg
CO2 Accuracy0 - 40mm ±2 mmHg
41 - 76 mmHg ±5% of reading
77 - 114 mmHg ±8% of reading
Above 80 BPM ±12% of reading
CO2 StabilityDrift   over   four   hours   shall   not   exceed   1  mmHg maximum
Accuracy specification will be maintained over a 120-hour  period
Sampling Frequency100 Hz
Respiration Rate Range2 to 150 Breaths Per Minute (BPM)
Respiration Rate Accuracy±1 breath
CompensationsAutomatic Barometric pressure 400 mmHg to 800 mmHg Operator selectable O2, N2O, He and agent compensation
CalibrationNo routine user calibration required.
Voltage Requirements5.0 VDC ±5%
Power ConsumptionLess   than   1.2   Watts   typical   (Steady  State)
Up to 2 Watts maximum on power up (Warm up)


Interconnection
Standard Pin out:
  1. VA 5.0V
  2. Shield Shield
  3. DGND Digital return
  4. VSRC 5.0V
  5. TxD Serial data from Sensor
  6. RxD Serial data from Host
  7. AGND Analog return
  8. SYNC Unused
Data InterfaceRS232, bi-directional, 19200 baud rate, standard N-8-1.
Data OutputCO2  gas concentration(mmHg), End-tidal CO2, Inspired CO2, Respiratory Rate.
Gas and barometric pressure compensated when supplied by host.
Temperature and HumidityOperating: 0 to 45ºC, 10 to 90% RH, non-condensing Storage: -40 to 70ºC, <90% RH, non-condensing
Water ResistanceIPX4 - Splash-proof (sensor head only)