Menu
Global B2B sourcing platform
Verified suppliers and manufacturers

High Precision Loop Resistance Test Unit Switch Contact Resistance Tester Multifunction Loop Impedance Tester

Price $1085.00
Min Order 1 sets
Availability In Stock
Shipping From Hubei, China
Popularity 384 people viewed
Model NO.:
KDHL-100P
After-sales Service:
Yes
Warranty:
1 Year
Display Type:
Digital Only
Testing Voltage:
220V
Operating Temperature:
10c~40c
Selling Units:
Single Item
Dimensions:
300*270*200 mm3
Accuracy Class:
0.5%
Weight:
5kg (Excluding Accessories)
Measurement Range:
1-1999
Measuring Range:
1~1999
Volume:
300*270*200mm3
Single Gross Weight:
13.000 Kg
Measurement Accuracy:
0.5%
Resistance Display Mode:
Three and a Half Position LCD
Test Current:
100A
Current Display Mode:
Three and a Half Position LCD
Working Power Supply:
AC220V10%.50Hz1Hz
HS Code:
9030332000
Product Description

Products introduction

The KDHL-100P circuit resistance tester can easily measure the contact resistance of the main contact of high and low voltage switches, the DC resistance of high and low voltage cable lines and other pure resistance test objects.

This product overcomes the disadvantage of traditional circuit resistance measuring instruments that require adjusting the current output; The large screen graphic LCD with wide temperature band backlight is adopted instead of the traditional Nixie tube display mode, which can display Chinese characters intuitively, making the operation of the instrument more convenient and simple. This product uses a precision instrument operational amplifier and a high-precision A/D converter, combined with a high-performance 16 bit microcontroller, to ensure the testing accuracy of the product;

This product has also designed functions such as real-time calendar clock, data storage, printing and protection, which makes the product more complete and cost-effective.

Environmental conditions for use

Power supply: AC220V ± 10%, 50Hz ± 1Hz
Operating temperature: 0C40C (no condensation)
Relative humidity: 90% RH
Altitude: less than 1000 meters

 
Measurement conditions
Environmental temperature(20 ± 5) C
Measurement of relative humiditynot exceeding 80%
There should be no external electromagnetic field that affects the normal operation of the instrument
 