Model NO.:
SKZ1052D
After-sales Service:
1 Year
Warranty:
1 Year
Type:
Differential Scanning Calorimeter
Display:
Digital
Control:
Computer Control
Weight:
0-100Kg
Power Source:
AC220V
Temperature Range:
-35-550c
Temperature Resolution:
0.1c
Heating Rate:
1-40c /Min
DSC Range:
600MW
Cooling Rate:
1 -20c / Min
Temperature Fluctuations:
0.1c
Temperature Repeatability:
0.1c
DSC Noise:
0.01um
DSC Resolution:
0.01um
DSC Accuracy:
0.01MW
Curve Scanning:
Rising Scan, Cooling Scan
Data Interface:
Standard USB Connector
Software:
Operating Software
Trademark:
SKZ
Transport Package:
Wooden Box Package
Specification:
68*62*55 42kg
Origin:
China
HS Code:
9027809990
Product Description
Product Description

Skz1052D Testing Equipment Simultaneous Tga Isothermal Oit DSC Differential Scanning Analysis Instrument Machine Analyzer
DSC is designed to determine the inner heat transition relating to temperature and heat flow, it is widely used in the field of polymer development, performance testing & quality control. DSC research and development includes the following field: glass transition temperature, melting point, cold crystallization, crystallization, phase transition, oxidation induction time (OIT).
Standards:
ISO/TR10837:1991, ASTM D3895-1998, ASTM E 967, ASTM E 968, ASTM E 793, ASTM D 3895, ASTM D 3417, ASTM D 3418, ISO 11357-6
Main Features:
1. New designed oven structure ensures high resolution and good stability of base line
2. Air flow meter may control the air flow rate accurately; the test data can be recorded

3. It is bilateral control, may be controlled by both main frame and software.

Technical parameters:
DSC range0~±500mW
Temperature range-35~550ºC    Compressor cooling
Heating rate1 ~ 40 ºC / min
Cooling rate1 ~ 20 ºC / min above room temperature,
1-10ºC / min, below room temperature.
Temperature resolution0.1 ºC
Temperature fluctuations±0.1ºC
Temperature repeatability±0.1ºC
DSC noise0.01μW
DSC resolution0.01μW
DSC accuracy0.1 u W
DSC sensitivity0.1 u W
Control ModeRising temperature, constant temperature(full automatic)
Curve scanningRising scan, cooling scan 
Gas flow controlEmbedded digital flow meter & Software control, automatic changing
Display 24 bit 7 inches LCD touch screen display
Data interfacestandard USB connector
Parameter standardWith standard material, with calibration function, the user may correct temperature and heat enthalpy