Percentage of liquid aspiration among total capacity | CV (Coefficient of variation) |
2% | <1.5% |
>10% | <0.5% |
A full remove for cylinder: 12.7mm

Description:
Piston pump & valve integrated module consists of Keyto's 2-way rocker diaphragm type solenoid valve and Keyto's precision piston pump.
It provides precise fluid dispensing for a variety of medical and laboratory instruments.
It reduces the leakage risk as less tubes and connectors installed, and has the high accuracy and consistency.
Specification:
| Max Dispense Volume | 100μL |
| Piston diameter (mm) | 3.17 |
| Resolution (μl/step) | 0.05 |
| Remove (mm) | 12.7 |
| Remove per step (mm/step) | 0.00635 |
| Motor rotatory angle per step (°) | 1.8 |
| CV (Coefficient of variation) | <1.2%; <1% (Measuring when more than 2%; 10% of total dispense) |
| Anticipated Life | 5 million cycles |
Connection Port | M6, 1/4-28 UNF |
| Integrated valve | match with 1013 series 2-way solenoid valve |
Mediums | Water, weak acid and base etc |
Operating Pressure(MPa) | 0.3 |
Note: 1: Please contact us if the medium is air. 2: The CV is tested under 2% range. 3: Standard precision is tested under 2% range. 4: It should be verified if using other mediums. More questions, please contact us. | |
Products photo



















Audited Supplier