CN6002 torque and speed measuring instrument, in order to enable users to measure and display field data more intuitively, the table provides the following functions:
Timing measurement at equal intervals, the measurement period T=20--2000ms can be set, and the timing pulse counting method is used to synchronously measure the torque and speed values at the current moment.
The display window provides three 5-digit display windows to display the measured values of torque, speed and power respectively, and can switch to display the maximum torque value in the measurement process, and give positive and negative torque indications.
The meter is connected to the sensor via a special five-core cable as follows:
Check whether the instrument power switch is off.
Connect the plug of one end of the special cable with the aviation socket on the rear panel of the instrument, pay attention to align the limit slot, insert it securely and then tighten it.
Connect the other end with the aviation socket of the sensor in the same way as above.
Turn on the instrument power can be measured

Zeroing: It has the function of calibrating the zero point of sensor measurement.
Range setting: the instrument provides the user input function of range setting, which can ensure the universality of the instrument and the sensor, and no special calibration is required.
Alarm: The user can set the alarm value of torque and speed, and the instrument has the function of alarm indication and provides alarm output signal.
Printing: It can be connected to an external printer for real-time data printing.
Communication: Provide standard RS-232 interface and communication protocol.
Power Detection Reset: Avoid crashes caused by power supply interference and fluctuations.
Parameter saving: You can save all kinds of set parameters.
The input and output are all using signal isolation technology, which has a certain anti-interference ability.
Technical Indicators:
Speed inaccuracy: ≤0.1%
Torque inaccuracy: ≤0.1%F•S
Insulation resistance: ≥500MΩ
Ambient temperature: -20~60°C
Ambient humidity: ≤90%RH
Power consumption: ≤15W