No. | Item | Specification |
| 1 | Magnet System | |
| 1.1 | Magnet Type | Permanent magnet with automatic constant temperature system |
| 1.2 | Field strength | 0.42T |
| 1.3 | Magnet shape | C-shape |
| 1.4 | Homogeneity (40cm, DSV, VRMS) | 2ppm |
| 1.5 | Shim method | Active/Passive/Dynamic |
| 1.6 | Magnet vertical gap | 40cm |
| 1.7 | Accessibility (Horizontal opening angle) | 250° |
| 1.8 | 5 Gauss fringe field | 2.5m*2.5m*2.5m |
| 2 | Gradient System | |
| 2.1 | Gradient field strength (Single Axis) | 31mT/m |
| 2.2 | Gradient slew rate (Single Axis) | 107mT/m/ms |
| 2.3 | Rise time | 0.29ms |
| 2.4 | Gradient cooling system(Gradient coils and power electronics) | Air |
| 3 | RF System | |
| 3.1 | Spectrometer | Fully digital |
| 3.2 | RF system type | Digital Transmit and Receive |
| 3.3 | Number of RF channel | 4 |
| 3.4 | Transmitter amplifier peak power | 6kW |
| 3.5 | RF bandwidth of receiver | Each 400KHz |
| 3.6 | Dynamic range | 80dB |
| 3.7 | Noise factor | 0.3dB |
| 3.8 | Coil type | phase array |
| 3.9 | Head coil | Yes |
| 3.10 | Neck coil | Yes |
| 3.11 | Body coil (small) | Yes |
| 3.12 | Body coil (large) | Yes |
| 3.13 | Knee coil | Yes |
| 3.14 | Ankle coil | Yes |
| 3.15 | Shoulder coil | Optional |
| 3.16 | Soft coil | Optional |
| 4 | Patient communication | |
| 4.1 | Max. patient weight | 210Kg |
| 4.2 | High accurate position assist | Yes |
| 4.3 | Localizer | Yes |
| 4.4 | Position accurate | 1mm |
| 4.5 | Position accessories | Yes |
| 4.6 | Emergence call | Yes |
| 4.7 | Communication system | Yes |
| 5 | Computer system | |
| 5.1 | Host computer | Dual core Industrial computer |
| 5.2 | MRI software | V30 |
| 5.3 | OS | WINDOWS 7 |
| 5.4 | CPU clock rate | 3.0GHz |
| 5.5 | Main memory | 4GB |
| 5.6 | Color LCD monitor | 24" |
| 5.7 | Keyboard and mouse | Yes |
| 5.8 | Hard disk | 500GB |
| 5.9 | Media driver | CRW / DVD |
| 5.10 | DICOM 3.0 | Yes |
| 5.11 | Ethernet | Yes |
| 6 | Scanning parameter | |
| 6.1 | Max. FOV | 400mm |
| 6.2 | Min. FOV | 10mm |
| 6.3 | Min. 2D Thickness | 1mm |
| 6.4 | Min. 3D thickness | 0.1mm |
| 6.5 | Max. Image matrix | 1024*1024 |
| 7 | Scanning sequence &imaging technology | |
| 7.1 | Spin echo 2D/3D (SE 2D/3D) | Yes |
| 7.2 | Multi-slice multi-echo(MSME) | Yes |
| 7.3 | Gradient echo(GRE 2D/3D) | Yes |
| 7.4 | Steady state process gradient echo( SSPGRE) | Yes |
| 7.5 | Fast spin echo (FSE) | Yes |
| 7.6 | Fast double echo(FDE) | Yes |
| 7.7 | Fast recover fast spin echo(FRFSE) | Yes |
| 7.8 | Single shot fast spin echo(SSFSE) | Yes |
| 7.9 | Multi shot fast spin echo (MSFSE) | Yes |
| 7.10 | Inversion recovery (IR) | Yes |
| 7.11 | Inversion recovery fast spin echo (IRFSE) | Yes |
| 7.12 | Short time inversion recovery (STIR) | Yes |
| 7.13 | Fluid attenuated inversion recovery (FLAIR) | Yes |
| 7.14 | Gating technology | Yes |
| 7.15 | Breath holding scan | Yes |
| 7.16 | TOF 2D MRA | Yes |
| 7.17 | TOF 3D MRA | Yes |
| 7.18 | Diffusion weighted imaging (DWI) | Yes |
| 7.19 | Max. b Value | 1000s/mm2 |
| 7.20 | Magnetization transfer (MTC) technology | Yes |
| 7.21 | MIP | Yes |
| 7.22 | MinIP | Yes |
| 7.23 | 3D position scan | Yes |
| 7.24 | Pre-saturation technology (PS) | Yes |
| 7.25 | Pre-saturation adjustment technology | Yes |
| 7.26 | Part metal implant scan technology | Yes |
| 7.27 | Automatic coil tuning | Yes |
| 7.28 | MRM | Yes |
| 7.29 | MRU | Yes |
| 7.30 | MRCP | Yes |
| 7.31 | Online image filtration | Yes |
| 7.32 | Online post procession | Yes |
| 7.33 | Optimal algorithm of active shimming, must be available to check the shimming data at any time | Yes |
| 7.34 | Scan sequence queuing | Yes |
| 7.35 | Multi-layer and multi-angle scanning technology | Yes |
| 7.36 | Optimize bandwidth acquisition technology | Yes |
| 7.37 | Flow compensation | Yes |
| 7.38 | Parallel acquisition technology | Yes |
| 7.39 | Section acquisition technology | Yes |
| 7.40 | Scan parameter preset | Yes |
| 7.41 | Oversampling technology | Yes |
| 7.42 | Image fusion technology | Yes |
| 7.43 | Artifact suppression technology | Yes |
| 7.44 | Thin imaging technology | Yes |
| 7.45 | Movie playback technology | Yes |
| 7.46 | Post processing package | Yes |











Audited Supplier