Percentage of liquid aspiration among total capacity | CV (Coefficient of variation) |
2% | <1.5% |
>10% | <0.5% |
Description:
Piston pump & valve integrated module consists of Keyto's 3-way rocker diaphragm type solenoid valve and Keyto's injection mould piston pump.
It provides precise fluid dispensing for a variety of medical and laboratory instruments.
It reduces the leakage risk as less tubes and connectors installed, and has the high accuracy and consistency.
Specification:
Max Dispense Volume | 1 mL |
Piston diameter | 10.01 mm |
Volume per stroke | 0.5 μL |
Travel distance | 12.7 mm |
Distance per step | 0.00635 mm |
Motor step angle | 1.8° |
CV (Coefficient of variation) | <1.2%; <1% (Measuring when more than 2%; 10% of total dispense) |
Anticipated Life | 5 million cycles |
Mounting | M6, 1/4-28 UNF |
Note: 1: Please contact us if the medium is air. 2: The CV is tested under 2% range. 3: Standard precision is tested under 2% range. 4: It should be verified if using other mediums. |
Product photo