Menu
Global B2B sourcing platform
Verified suppliers and manufacturers

GF302 Portable High Precision Multi-Product Calibrator

Price $5500.00
Min Order 1 Piece
Availability In Stock
Shipping From Beijing, China
Popularity 421 people viewed
Model NO.:
GF302
Warranty:
Three Years
Application:
Mechanical Instrument
Power:
Electricity
Certification:
FDA, CE, IAF, CCC
Voltage Output:
50V, 100V, 200V, 400V, 600V
Current Output:
0.5A, 1A, 2.5A, 5A, 10A, 20A
LCD Display:
With LCD Display
Customized:
Customized
Keyword:
Portable High Precision Multi-Product Calibrator
Model:
Multi-Product Calibrator
GF302:
Multi-Product Calibrator High Precision
Accuracy Class:
0.05%, 0.1%
Communication Port:
USB, RS232
Load Capacity:
25va
Phase Angle Scope:
0-359.99
Weight (Kg):
20
Trademark:
GFUVE
Transport Package:
Standard Export Wooden Box
Specification:
460x430x185
Origin:
Beijing, China
Product Description
GF302 Portable High Precision Multi-Product Calibrator
Product Description

 


GF302 portable multifunction instrument calibrator is suitable for power plant and power grid companies for the following function: measuring and testing department and instrumentation classes, national levels measuring and testing institutions, railway, petroleum, chemical industry and other large industrial and mining enterprises, scientific research units, etc. The core technology function with digital signal processor (DSP) and 16 high-speed digital converters composed of high precision work frequency communication terminal. The signal source is DSP and 16 high-speed digital-to-analog converters, it can control the sine wave and distortion wave signal source.
Functions and Features
All kinds of electric measurement transducer can be checked, includingAC/DC voltage transducer, AC/DC current transducer, frequency transducer, phase transducer, single/ three-phaseAC active power transducer, three-phase reactive power transducers
Check all kinds of electric measurement indicating meter, includingAC/DC voltmeter, AC/DC ammeter, frequency meter, phase meter, single three-phase ac active power meter, three-phase ac reactive power meter, synchronous meter, etc
Test single-phase, three-phase electronic, mechanical watt-hour meter or energy meter/ kWh meter error
Calibrate AC sample device, RTU, measurement device error
The built-in electric measurement transducer, electric measurement instrument and meter instructions of verification procedures, can fully automatic or semi-automatic for verification, and save 1000 group check data
It can be used as voltage source, current source and power source with high precision, and it isa high stability standard resource
6.4-inch big screen color display and English interface
For the software calibration, you don't need to open the case, it's stable and reliable
Voltage output terminal with short circuit, current output terminal open protection and power amplifier overheating protection function
With automatic failure detection function, shows fault part, the convenience users check line
With USB port, it can connect computer for data management or controlled by PC
Product Parameters

 

Electrical parameters
Accuracy class0.05%, 0.1%
Power supplySingle phase AC 220V±10% or 110V±10%, 50/60Hz
Communication portUSB, RS232
AC Voltage output
Range(U1,U2,U3)50V, 100V, 200V, 400V, 600V
Adjusting range(0-120)% RG
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01% / 1 min
Load capacity25VA
Output distortion degree≤0.3% or (linear load)
AC Current output
Range(I1,I2,I3)0.5A, 1A, 2.5A, 5A, 10A, 20A
Adjusting range(0-120)% RG
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01%/1 min
Load capacity25VA
Output distortion degree≤0.3% or (linear load)
AC Power output
Accuracy0.05% RG
Stability0.01%/1min
Frequency
Frequency range45.000 - 65.000 Hz
Resolution0.001 Hz
Accuracy0.002 Hz
Power factor output
Adjusting range-1 ~ 0 ~ 1
Adjust fineness0.0001
Accuracy0.0005
Phase angle
Scope0°-359.99°
Resolution0.01°
Accuracy0.05°
Voltage/Current harmonic output
Times2-31st
Content0-40%
Phase0-359.999 degree
Configuration error(10% RD + 0.1%), RD refers to the configuration value of harmonic contents
DC Voltage output
Range75mV, 75V, 150V, 300V, 500V, 1000V
Adjusting range(0-120)% RG
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01%/1min
Load capacity25VA
DC Current output
Range0.5A, 1A, 2.5A, 5A, 10A, 20A
Adjusting range(0-120)% RG
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01%/1min
Load capacity25VA
DC measurements
DC voltage measurement range0 to ±24V
DC current measurement range0 to ±24mA
Measurement accuracy0.01% RG
Watt-hour meter measuring the integrated error
Active energy0.05%
Reactive energy0.1%
Mechanical parameters
Dimensions (W×H×D) (mm)460x430x185
Weight (kg)20
Environmental conditions
Working temperature0°C to 40°C
Storage conditions-30°C to 60°C
Relative humidity≤85%
  • Similar Products