Model NO.:
GF302C
After-sales Service:
24 Hours Service
Warranty:
3 Years Warranty
Application:
Mechanical Instrument, Electromagnetic Instrument
Power:
Electricity
Certification:
FDA, CE, IAF, CCC
Voltage Output:
0-720V AC/DC
Current Output:
0-24A AC/DC
Thermocouple (TC):
No
Resistance-Temperature Detector (RTD):
No
LCD Display:
With LCD Display
Customized:
Customized
Accuracy:
0.05%
Weight:
18kg
Size:
460mmx430mmx185mm
Harmonic:
2~31
DC Output Capacity:
30W
Function 1:
Test Power Meter
Function 2:
Test AC/DC Ammeter
Function 3:
Test AC/DC Voltmeter
Function 4:
Test Frequency Meter
Function 5:
Test Phase Meter
Trademark:
GFUVE
Transport Package:
Standard Export Carton
Specification:
CE, ISO9001: 2008, IEC
Origin:
Beijing, China
HS Code:
9030849000
Product Description
GF302C PORTABLE PANEL METER CALIBRATOR WITH AC/DC VOLTAGE AND CURRENT SOURCE

This device is according to nation verification regulation JJG124-2005 "ammeter, voltmeter, power meter and resistance meter calibration regulations and the relevant countries standard requirements and design three-phase 0.05 magnitude meter source integration calibration device". The core technology of this device is with digital signal processor (DSP) and 16 high-speed digital converter, which is composed of high precision work frequency communication terminal; The signal source use the DSP and 16 high-speed digital-to-analog converters, which can control the sine wave, distortion wave signal source; Equipment has high precision, stable and reliable, and easy to operate flexible, and other characteristics; Electric power system is used for electrical measurement is the ideal calibration equipment.
 

FEATURES
It can check all kinds of electric measurement of instrument including: AC/DC voltmeter, AC/DC ammeter, frequency meter, phase angle meter, single/three-phase AC active power meter, single/three-phase AC reactive power meter, synchronous meter, etc.;
Built-in electric measurement of verification procedure indicating meter, automatic or semi-automatic for verification, saving 1000 groups data;
It can as a voltage source, a current source and power source for high precision and high stability standard resource;
6.4 inch big screen color display;
It is calibrated in the software and don't need to open the case, stable and reliable;
Having short circuit, current output terminal, output voltage open protection and power amplifier overheating protection function;
Having automatically failure detection function, shows fault part, it is convenience for users to check line;
With USB port and PC connection for data management.
 APPLICATION
It is suitable for power plant and substation work area of grid companies, test division and the instrument, and national levels measuring and testing institutions, such as railway, petroleum, chemical industry and other large industrial and mining enterprises, scientific research units, etc.
PARAMETERS

Electrical parameters
Accuracy class 0.05%, 0.1%
Power supplySingle phase AC 220V±10% or 110V±10%, 50/60 Hz 
or AC 110 V±10%, 50/60 Hz
Communication Port USB port
AC voltage output
Range(U1,U2,U3)50 V, 100 V, 200 V, 400 V, 600 V
Adjusting range(0-120) % RG
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01%/1min
Load capacity25 VA
Output distortion degree≤0.3% (linear load)
AC current output
Range(I1,I2,I3)0.5A, 1A, 2.5A, 5A, 10A, 20A
Adjusting range(0-120) % RG
Adjust fineness0.005% RG 
Accuracy0.05% RG
Stability0.01%/1min
Load capacity25 VA 
Output distortion degree≤0.3% (linear load) 
DC voltage output
Range75 V, 150 V, 300 V, 500 V, 1000 V
Adjusting range(0-120) % RG 
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01%/1min 
Load capacity25 VA 
DC current output
Range0.5A, 1A, 2.5A, 5A, 10A, 20A 
Adjusting range(0-120)% RG
Adjust fineness0.005% RG
Accuracy0.05% RG
Stability0.01%/1min 
Load capacity25 VA
AC power output 
Accuracy0.05% RG
Stability0.01%/1min 
Frequency
Frequency range45-65 Hz
Resolution0.001 Hz
Accuracy0.002 Hz
Power factor output
Adjusting range-1 to 0 to +1 
Adjust fineness0.0001
Accuracy0.0005
Phase angle
Scope0°-359.99°
Resolution0.01°
Accuracy0.05° 
Voltage/Current harmonic output
Times2nd-31th
Content0-40%
Phase0°-359.999°
Configuration error(10% RD + 0.1%), RD refers to the configuration value of harmonic contents
Mechanical parameters
Dimensions (W×D×H) (mm)460x430x185
Weight (kg)18
Environmental conditions
Working temperature0°C to 40°C
Storage conditions-30°C to 60°C
Relative humidity≤85%