TA1500 AC / DC Standard Source

Price $420000.00 Compare
Min Order 1 set
Availability In Stock
Shipping From Hunan, China
Popularity 235 people viewed
Quantity
+-
 

Tunkia Co., Ltd.

VIP   Audited Supplier 4 years
Profile Certified by SGS/BV
TA1500
Life Time
12 Months Excellent After-Sales Service
Electricity
6ka
Non-Customized
TUNKIA
Package for Exporting
China
Product Description
TA1500 AC Direct Current Standard Source 
Schematic diagram of a 1kA specification current source
*Reference only.

1. Summary

TA1500 is a series of AC current standard source devices, which can output high stability and high accuracy AC current large current, suitable for calibrating AC current ammeters, current transformers, Current sensors and other devices.
 

2. Features

  1. Supports combined outputs from multiple module sources up to 6 kA.
  2. Test frequency: DC, 10 Hz ~ 65 Hz (400, 1kHz optional).
  3. Accuracy class: 0.02 or 0.05
  4. AC and DC current outputs use the same bus and terminal block
  5. DC current is bipolar and does not require relay commutation.
  6. Support external control console to adjust the current output.
  7. The software controls the slow rise and fall of the large current, and the overshoot amplitude is small.
 

3. Applications

 Calibrate the AC current ammeter
  1. Class 0.02 specification: Suitable for calibrating large AC-DC ammeter of class 0.1 and below.
  2. Class 0.05 specification: Suitable for calibrating large AC-DC ammeter of class 0.2 and below.
 

☆ Calibrate current sensor / transformer
  1. TA1500 can be used as a primary current input to the sensor or transformer under test.
  2. In combination with a precision ammeter or voltmeter, the sensor or transformer is calibrated using the direct measurement method.

4.  Functional features

☆ Measurement output and adjustment mode
  1. With the "fixed-point output" method, you can directly set the required output value by clicking the touch screen.
  2. The unit switches fully automatically to the optimal range output.
 
☆ Harmonic output (optional).
  1. Optional harmonic output function, current can be superimposed 2~21 harmonics @ 50 Hz.
  2. The amplitude and phase of each harmonic can be set independently, and the output waveform and spectrogram can be displayed.

5. Specifications

5.1 AC direct current output

RangeShort-term stability
(%/min)
Measurement uncertainty (k=2)
(ppm*RD+ppm*RG) [1]
Maximum load
Voltage (Vpk).
Class 0.01 [2]Class 0.02Class 0.05 Class 0.01 [2]Class 0.02 Class 0.05
10 A0.0030.0050.0160 + 40150 + 50400 + 1008
30 A0.0030.0050.0160 + 40150 + 50400 + 1008
100 A0.0030.0050.0160 + 40150 + 50400 + 1008
300 A0.0030.0050.0160 + 40150 + 50400 + 1008
600 A0.0030.0050.0160 + 40150 + 50400 + 1008
1 kA0.0030.0050.0160 + 40150 + 50400 + 1008
························
N*1 kA0.0030.0050.0160 + 40150 + 50400 + 1008
Note: [1] RD is the reading value, RG is the range value;  
[2] The 0.01 level only supports DC, 45 ~ 65 Hz frequency range.
  1. Output range: range below 1 kA (excluding 1 kA):  10%~112%*RG
Range above 1 kA (including 1 kA): 10%~100%*RG (maximum expandable output to 6 kA )
  1. Fineness adjustment: 5 ppm*RG, 7-digit decimal display
  2. DC ripple coefficient: ≤ 0.5% @ 5 kHz or less, AC distortion: ≤ 0.5%.
  3. Settling time: Using the slow rise output mode, the time to stabilize to the output accuracy is about 3 s
  4. Protection function: open circuit protection, overload protection, overheating protection
 

5.2 Frequency and Harmonics

Sine wave frequencyFrequency range10 Hz ~ 65 Hz(400 Hz,1 k Hz optional)
Minimum fineness0.0001 Hz
Measurement uncertainty (k=2).0.005%
Harmonic output
(Optional)
Number of harmonics2nd ~ 21st time
Harmonic amplitude0.0% ~ 40.0% adjustable (relative to fundamental).
Measurement uncertainty (k=2).0.5%

6. General Specifications

Power supplyAC (380 ± 38) V,(50±2) Hz
Maximum power consumption16 kVA / per1 kA
Warm-up time30 minutes
Temperature performanceWorking temperature: 10°C~35°C
Calibration temperature: 18°C~28°C
Storage temperature: -20 °C ~ 70 °C
Humidity performanceOperating humidity: <80% @ 30°C, <70% @ 40°C
Storage humidity: (20%~80%) R· H, no condensation
altitude< 3000 m
Communication interfaceRS232 interface
 

7. Ordering Information



 
Company Profile
 





  • Contact Supplier
Message Type
* Message Content
* Name  
* Phone
* Email
* CAPTCHA  
$999.00 1 Piece(MOQ)

TD1600 AC Source for AC Current Meter Voltage Meter Calibration

Tunkia Co., Ltd.
$30000.00 1 Set(MOQ)

TUNKIA TD1330 Portable AC EV charging station tester charging pile tester

Tunkia Co., Ltd.
Negotiable 1 set(MOQ)

TH1300 DC High Current Shunt with Sensor Calibration System

Tunkia Co., Ltd.
$3650.00 1 Piece(MOQ)

Handheld digital LCD static electricity leakage Surface Resistance meter Tester Surface Insulation Resistance meter (SIR Testing)

Tunkia Co., Ltd.
$100000.00 1 Set(MOQ)

TUNKIA TK4830 AC DC EV Charger tester Portable Tester for EV Charging Stations

Tunkia Co., Ltd.
$14000.00 1 Set(MOQ)

TD1858 Portable Multifunction Calibrator for 4.5 Digit Multimeter Testing

Tunkia Co., Ltd.
$999.00 1 Piece(MOQ)

TD1600 AC Source for AC Current Meter Voltage Meter Calibration

Tunkia Co., Ltd.
$880000.00 1 Set(MOQ)

High Precisous Small Caliber Volumetric Method Industrial Use Calibration Equipment Syestem of Liquid Flow

Shanghai Powerbuilder Automation Technology Co., Ltd.
$420000.00 1 set(MOQ)

TA1500 AC / DC Standard Source

Tunkia Co., Ltd.
$100.00 2 Pieces(MOQ)

Calibration Methods Standard Calibration Multifunction Process Calibrator

Hangzhou Supmea International Trading Co., Ltd.